为评估室内LED光源潜在的光生物安全风险,搭建了光生物安全测试系统平台.在该系统平台上对常用的几款室内LED光源进行了较全面的光生物安全测试.并对照标准GB /T 20145—2006《光源和光源系统的光生物安全》的要求,对实验结果作了较为详细的分析,对该光源的危险类别做了判定划分.该报告分析对搭建室内LED产品的光生物安全性测试系统以及研究其测试方法均具有一定的参考意义.
Testing platform for photo-biological safety was built in our laboratory in order to evaluate the potential photo-biological safety risk of indoor LED light sources. Several kinds of indoor LED light sources which were extensively used were tested by this system, mainly including the testing of irradiance and radiance. Furthermore, the work analyzed the experimental results according to the requirement of the standard GB /T 20145—2006: photo-biological safety of lamps and lamp system. At last, the hazard class of the light sources was evaluated and determined. The results have a certain reference significance for the research of testing system and testing method for photo-biological safety of the LED products.
[1] HELIOTOTIS G, STSVNNOUS P N, BRADLEY D D C. Spectral conversion of InGaN ultraviolet microarray light emitting diodes using fluorine-based red, green, blue and white light emitting polymer over layer films[J]. Applied Physic Letters, 2005,87: 503-504.
[2] 俞安琪,赵旭.LED照明产品光生物安全风险状况分析[J].照明工程学报,2013,24(5):1-6.
[3] 蔡怡,朱腾飞,张俊.一种小型光色度计的设计[J].光学仪器,2012,34(3): 27-32.
[4] 赵浩之,蒋春旭,王深,等. 基于EN62471:2008的LED灯串光生物安全检测与分析[J]. 照明工程学报,2013,24(S1):115-121.
[5] 虞建栋,王建平.LED光辐射安全及标准进展[J].信息技术与标准化,2009(5):14-16.
[6] 陈慧挺,蔡喆,吴晓晨,等. 大功率LED路灯的光生物安全测试与分析[J]. 照明工程学报,2011,22(6): 88-90. DOI:10.13223/j.cnki.ciej.2011.06.001.
[7] JJF(浙)1146—2018,直流高压试验装置校准规范(浙江省地方计量技术规范)[S].
[8] 蔡怡,朱腾飞,汪哲弘,等. LED照明产品的光生物安全测量的不确定度评定[J]. 光学仪器 2017,39(4): 13-17.
[9] GB/T20145—2006,光源和光源系统的光生物安全性[S].
[10] 刘艳珍,孟祥峰,李宁,等. 医疗器械非相干光辐射视网膜热危害研究[J]. 中国医疗设备,2017,32(9): 10-14. (